RecallRadar

FDA Device recall Z-1840-2019

LEEP Precision Integrated System 120V, Model LP-10-120 Product Usage: The LEEP PRECISION INTEGRATED SYSTEM l 20V combines three components…

On 2019-06-26 the FDA classified a Class II recall of LEEP Precision Integrated System 120V, Model LP-10-120 Product Usage: The LEEP PRECISION INTEGRATED SYSTEM l 20V combines three components… by Coopersurgical, citing 2.5A slow blow fuses may have been installed in the LEEP RPECISION Generator rather than the appropriate 3.15A slow blow fuses.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1840-2019
ClassificationClass II
Statusterminated
Initiated2019-04-23
Published2019-06-26
Last updated2026-09-13 11:55 UTC
CompanyCoopersurgical
DistributionUS

Product

LEEP Precision Integrated System 120V, Model LP-10-120 Product Usage: The LEEP PRECISION INTEGRATED SYSTEM l 20V combines three components - LEEP PRECISION Cart, LEEP PRECISION Generator, and the LEEP PRECISION Smoke Evacuator for loop electrosurgical excision procedures. The loop electrosurgical excision procedure (LEEP) utilizes a very thin wire in the shape of a loop heated by electrical current generated by a power supply to efficiently remove abnormal cells. The LEEP PRECISION INTEGRATED SYSTEM has been designed to provide an efficient workstation that can efficiently meet all the unique LEEP requirements of the clinic and practice-based facilities .

Reason

2.5A slow blow fuses may have been installed in the LEEP RPECISION Generator rather than the appropriate 3.15A slow blow fuses.

Identifiers

code_info
2018110001,2018110002, 2018110003, 2018110004, 2018110005,2018110006, 2018110007, 2018110008,2018110011, 2018110012,201…2018110001,2018110002, 2018110003, 2018110004, 2018110005,2018110006, 2018110007, 2018110008,2018110011, 2018110012,2018110013, 2018110014, 2018110015,2018110016, 2018110017, 2018110019, 2018110020,2018110021,& 2018110022

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1840-2019%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.