RecallRadar

FDA Device recall Z-1837-2022

Roadrunner PC Hydrophilic Wire Guide, Hydrophilic Guidewire Reference Part Number/GPN: RPC-035145-0-5 G34131 RPC-035145 G18154 RPC-035145-5…

On 2022-10-05 the FDA classified a Class II recall of Roadrunner PC Hydrophilic Wire Guide, Hydrophilic Guidewire Reference Part Number/GPN: RPC-035145-0-5 G34131 RPC-035145 G18154 RPC-035145-5… by Cook, citing cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices ma….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1837-2022
ClassificationClass II
Statusongoing
Initiated2022-08-16
Published2022-10-05
Last updated2026-09-13 11:56 UTC
CompanyCook
DistributionUS

Product

Roadrunner PC Hydrophilic Wire Guide, Hydrophilic Guidewire Reference Part Number/GPN: RPC-035145-0-5 G34131 RPC-035145 G18154 RPC-035145-5 G34132 RPC-038145-0-5 G34129 RFSPC-038145-0 G17539 RFSPC-035145 G17540 RPC-038145-0 G18155 RPC-035145-0 G18153 RFSPC-035145-0 G17542

Reason

Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

Identifiers

code_info
RPC-035145-0-5 G34131 UDI-DI: 00827002341319 Lot/Expiration Date: 14769270X 03-06-2025 14802181 20-06-2025 14816224 28-…RPC-035145-0-5 G34131 UDI-DI: 00827002341319 Lot/Expiration Date: 14769270X 03-06-2025 14802181 20-06-2025 14816224 28-06-2025 14818488 29-06-2025 14818489 29-06-2025 14828627 06-07-2025 14828629 06-07-2025 RPC-035145 G18154 UDI-DI: 00827002181540 Lot/Expiration Date: 14770863 06-06-2025 14770864 06-06-2025 14813490 27-06-2025 14816227 28-06-2025 NS14770866 06-06-2025 NS14804982 21-06-2025 RPC-035145-5 G34132 UDI-DI: 00827002341326 Lot/Expiration Date: 14770872 06-06-2025 14810022 23-06-2025 14831440 07-07-2025 RPC-038145-0-5 G34129 UDI-DI: 00827002341296 Lot/Expiration Date: 14804972 21-06-2025 14809971 23-06-2025 RFSPC-038145-0 G17539 UDI-DI: 00827002175396 Lot/Expiration Date: 14809157 23-06-2025 14837048 12-07-2025 RFSPC-035145 G17540 UDI-DI: 00827002175402 Lot/Expiration Date: 14818484 29-06-2025 14818485 29-06-2025 14818490 29-06-2025 14840310 13-07-2025 RPC-038145-0 G18155 UDI-DI: 00827002181557 Lot/Expiration Date: 14819269 29-06-2025 14823406 01-07-2025 14832976 08-07-20

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1837-2022%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.