FDA Device recall Z-1831-2022
Double Flexible Tipped Wire Guide, Guidewire Reference Part Number/GPN TDOC-35-145-0-3 G00362 C-TDOC-18-50-0-2 G05146 TDOC-35-50-0-3 G00364…
- FDA Device
- Class II
- ongoing
- Published 2022-10-05
On 2022-10-05 the FDA classified a Class II recall of Double Flexible Tipped Wire Guide, Guidewire Reference Part Number/GPN TDOC-35-145-0-3 G00362 C-TDOC-18-50-0-2 G05146 TDOC-35-50-0-3 G00364… by Cook, citing cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices ma….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1831-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-08-16 |
| Published | 2022-10-05 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Cook |
| Distribution | US |
Product
Double Flexible Tipped Wire Guide, Guidewire Reference Part Number/GPN TDOC-35-145-0-3 G00362 C-TDOC-18-50-0-2 G05146 TDOC-35-50-0-3 G00364 C-DOC-21-40-0-2 G02431 C-TDOC-21-50-0-2 G05195 C-TDOC-25-50-0-2 G05145
Reason
Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
Identifiers
| code_info | TDOC-35-145-0-3 G00362 UDI-DI: 00827002003620 Lots/Expiration Date: 14793053 14-06-2027 14797769 16-06-2027 14802346 20…TDOC-35-145-0-3 G00362 UDI-DI: 00827002003620 Lots/Expiration Date: 14793053 14-06-2027 14797769 16-06-2027 14802346 20-06-2027 14802348 20-06-2027 14812034 24-06-2027 C-TDOC-18-50-0-2 G05146 UDI-DI: 00827002051461 Lots/Expiration Date: 14795065 15-06-2027 TDOC-35-50-0-3 G00364 UDI-DI: 00827002003644 Lots/Expiration Date: 14802347 20-06-2027 C-DOC-21-40-0-2 G02431 UDI-DI: 00827002024311 Lots/Expiration Date: NS14795068 15-06-2027 C-TDOC-21-50-0-2 G05195 UDI-DI:00827002051959 Lots/Expiration Date: NS14802356 20-06-2027 C-TDOC-25-50-0-2 G05145 UDI-DI: 00827002051454 Lots/Expiration Date: NS14818384 29-06-2027 NS14818387 29-06-2027 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1831-2022%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.