FDA Device recall Z-1831-2012
The product is Pathfinder SL, version 1.6.0.3057 software used with the Spacelabs Healthcare Pathfinder SL Holter Analyzer, version 1.6.0.…
- FDA Device
- Class II
- terminated
- Published 2012-06-27
On 2012-06-27 the FDA classified a Class II recall of The product is Pathfinder SL, version 1.6.0.3057 software used with the Spacelabs Healthcare Pathfinder SL Holter Analyzer, version 1.6.0.… by Spacelabs Healthcare, citing the firm received reports of the Pathfinder LS software crashing after editing a batch of Holter Analyzer records. The software will stop with an error message requiring a restart….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1831-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-05-11 |
| Published | 2012-06-27 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Spacelabs Healthcare |
| Distribution | US |
Product
The product is Pathfinder SL, version 1.6.0.3057 software used with the Spacelabs Healthcare Pathfinder SL Holter Analyzer, version 1.6.0. Product Usage: The Holter Analyzer analyzes information collected from ambulatory electrocardiogram recorders worn by patients and used to assist physicians in diagnosis and patient monitoring. In addition, incorrect QTc values are reported after moving the cursor. QTc values are timing events between labeled points of interest in the ECG signal that the physician can mark with the cursor. The software is used to edit a family of events and upon conclusion the software crashes with an error message.
Reason
The firm received reports of the Pathfinder LS software crashing after editing a batch of Holter Analyzer records. The software will stop with an error message requiring a restart. In addition, QTc values will not update after moving the marker.
Identifiers
| code_info | Version 1.6.0. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1831-2012%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.