FDA Device recall Z-1821-2017
UNIQ: UNIQ FD10 722026
- FDA Device
- Class II
- terminated
- Published 2017-04-26
On 2017-04-26 the FDA classified a Class II recall of UNIQ: UNIQ FD10 722026 by Philips Electronics North America, citing due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1821-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-03-22 |
| Published | 2017-04-26 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Philips Electronics North America |
| Distribution | US |
Product
UNIQ: UNIQ FD10 722026; UNIQ FD10/10 722027; UNIQ FD20/10 biplane 722029; UNIQ FD10 OR Table 722033; UNIQ FD10/10 OR Table 722034; UNIQ FD20/10 OR Table 722036; UNIQ FD20/20 722038; UNIQ FD20/20 biplane OR Table 722039
Reason
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
Identifiers
| code_info | Devices with a Laird Chiller. |
|---|