RecallRadar

FDA Device recall Z-1820-2018

Endo GIA" Auto Suture" Universal Articulating Loading Unit 60mm - 3.5mm ( Item Code 030458) Product Usage: The Endo GIA universal and Endo…

On 2018-05-23 the FDA classified a Class II recall of Endo GIA" Auto Suture" Universal Articulating Loading Unit 60mm - 3.5mm ( Item Code 030458) Product Usage: The Endo GIA universal and Endo… by Covidien Medtronic, citing device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading t….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1820-2018
ClassificationClass II
Statusterminated
Initiated2018-04-26
Published2018-05-23
Last updated2026-09-13 11:54 UTC
CompanyCovidien Medtronic
DistributionUS

Product

Endo GIA" Auto Suture" Universal Articulating Loading Unit 60mm - 3.5mm ( Item Code 030458) Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, triple-staggered rows of titanium staples and simultaneously divides the tissue between the two, triple-staggered rows. The size of the staples is determined by the selection of the 2.0 mm, 2.5 mm, 3.5 mm or 4.8 mm single use loading unit. The Endo GIA universal staplers will accommodate any of the single use loading unit sizes that are available in the 30 mm, 45 mm and 60 mm lines.

Reason

Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.

Identifiers

code_info
Lots: N7L0911KX, N7M0067KX, N7F0631KX, N7L0465KX, N7L0117KX, N7L0768KX, N7M0314KX,N7M0314KX,N7L1075KX, N7M0646KX, N7M07…Lots: N7L0911KX, N7M0067KX, N7F0631KX, N7L0465KX, N7L0117KX, N7L0768KX, N7M0314KX,N7M0314KX,N7L1075KX, N7M0646KX, N7M0740KX, N7M0972KX, N8A0172KX, N8A0539KX, N8A0679KX, N8A0910KX, N8A0961KX, N8B0054KX, N8B0099KX, N8C0008KX, N8C0205KX, N8C0532KX, N8C0571KX, N8C0758KX, N8C0795KX, N8C1284KRX GTIN # 1 (Inner Carton) - 10884523002898 GTIN # 2 (Outer Carton)- 20884523002895

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1820-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.