FDA Device recall Z-1818-2012

Philips Digital Diagnost Software, software release 2.0.2 and 2.0.2SP1 including mirror icon Stationary Fluoroscopic system

On 2012-06-27 the FDA classified a Class II recall of Philips Digital Diagnost Software, software release 2.0.2 and 2.0.2SP1 including mirror icon Stationary Fluoroscopic system by Philips Healthcare, citing with Digital Diagnost (DIDI) R 2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back i….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1818-2012
ClassificationClass II
Statusterminated
Initiated2012-05-22
Published2012-06-27
Last updated2026-09-13 11:53 UTC
CompanyPhilips Healthcare
DistributionUS

Product

Philips Digital Diagnost Software, software release 2.0.2 and 2.0.2SP1 including mirror icon Stationary Fluoroscopic system

Reason

With Digital Diagnost (DIDI) R 2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a circle" appears in the lower left corner. This can be mistaken for a "right patient side" marker, although this can appear on the left patient side

Identifiers

code_infoSoftware release 2.0.2 and 2.0.2SP1 including mirror icon

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1818-2012%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.