On 2020-05-06 the FDA classified a Class II recall of Q-Stress/XScribe by Welch Allyn Mortara, citing impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1817-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-03-18 |
| Published | 2020-05-06 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Welch Allyn Mortara |
| Distribution | US |
Product
Q-Stress/XScribe. Q-Stress or XScribe is a diagnostic device capable of real time ECG display, heart rate measurement, ST analysis and ventricular ectopic beat detection using wired or wireless acquisition modules.
Reason
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
Identifiers
| code_info | Model Numbers: QS-6XX-XXXXX, XSCRIBE-5XX-XXXXX, XSCRIBE-6XX-XXXXX. |
|---|