RecallRadar

FDA Device recall Z-1817-2020

Q-Stress/XScribe

On 2020-05-06 the FDA classified a Class II recall of Q-Stress/XScribe by Welch Allyn Mortara, citing impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1817-2020
ClassificationClass II
Statusterminated
Initiated2020-03-18
Published2020-05-06
Last updated2026-09-13 11:55 UTC
CompanyWelch Allyn Mortara
DistributionUS

Product

Q-Stress/XScribe. Q-Stress or XScribe is a diagnostic device capable of real time ECG display, heart rate measurement, ST analysis and ventricular ectopic beat detection using wired or wireless acquisition modules.

Reason

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Identifiers

code_infoModel Numbers: QS-6XX-XXXXX, XSCRIBE-5XX-XXXXX, XSCRIBE-6XX-XXXXX.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1817-2020%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.