FDA Device recall Z-1815-2020

S4 and patient cable for the S4 Wireless Telemeter with the following device names: CABLE 10 WIRE LF ECG SNAP ENDS AHA GRAY, CABLE 10 WIRE…

On 2020-05-06 the FDA classified a Class II recall of S4 and patient cable for the S4 Wireless Telemeter with the following device names: CABLE 10 WIRE LF ECG SNAP ENDS AHA GRAY, CABLE 10 WIRE… by Welch Allyn Mortara, citing impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1815-2020
ClassificationClass II
Statusterminated
Initiated2020-03-18
Published2020-05-06
Last updated2026-09-13 11:55 UTC
CompanyWelch Allyn Mortara
DistributionUS

Product

S4 and patient cable for the S4 Wireless Telemeter with the following device names: CABLE 10 WIRE LF ECG SNAP ENDS AHA GRAY, CABLE 10 WIRE LF ECG SNAP ENDS IEC GRAY, CBL 10WIRE LF LRG ECG SNAP ENDS AHA GRAY, CBL 10WIRE LF LRG ECG SNAP ENDS IEC GRAY. The S4 mobile monitor provides a means to acquire and transmit simultaneous ECG and SpO2 data to a Surveyor Central Station monitoring system while allowing the patient to be ambulatory within the range of the WiFi network.

Reason

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Identifiers

code_infoModel Numbers: S4-Q-ASX-XXX, S4-Q-AXX-XXX, S4-P-A, 9293-061-50, 9293-061-51, 9293-061-52, 9293-061-53.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.