FDA Device recall Z-1814-2025
Adult Manual Resuscitator with Medium Adult Mask, Bag REservoir, Filter, Manometer and 7 ft Oxygen Tubing, REF CPRM1116F
- FDA Device
- Class I
- ongoing
- Published 2025-06-11
On 2025-06-11 the FDA classified a Class I recall of Adult Manual Resuscitator with Medium Adult Mask, Bag REservoir, Filter, Manometer and 7 ft Oxygen Tubing, REF CPRM1116F by Sunmed Holdings, citing affected lots were manufactured with B/V Filter incorrectly attached to the wrong port (patient port instead of the exhalation port). If not noticed prior to patient use, there wo….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1814-2025 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2025-05-01 |
| Published | 2025-06-11 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Sunmed Holdings |
| Distribution | US |
Product
Adult Manual Resuscitator with Medium Adult Mask, Bag REservoir, Filter, Manometer and 7 ft Oxygen Tubing, REF CPRM1116F
Reason
Affected lots were manufactured with B/V Filter incorrectly attached to the wrong port (patient port instead of the exhalation port). If not noticed prior to patient use, there would be interruption or delay in patient resuscitation, which may lead to life threatening consequences, including hypoxia, hypercapnia, organ failure, and death.
Identifiers
| code_info | UDI-DI: EA: 10884389164822 CS: 40884389164823 Lots: 526782 526790 526796 526797 526798 526800 526802 526804 526805 5…UDI-DI: EA: 10884389164822 CS: 40884389164823 Lots: 526782 526790 526796 526797 526798 526800 526802 526804 526805 526806 526807 526808 526809 526810 526814 526815 526816 526811 526817 526818 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1814-2025%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.