RecallRadar

FDA Device recall Z-1812-2020

H12+ and patient cable for the H12+ Holter Recorders with the following device name: PAT CBL 10WIRE LEADFORM AHA SNAP, PAT CBL 10WIRE LEADF…

On 2020-05-06 the FDA classified a Class II recall of H12+ and patient cable for the H12+ Holter Recorders with the following device name: PAT CBL 10WIRE LEADFORM AHA SNAP, PAT CBL 10WIRE LEADF… by Welch Allyn Mortara, citing impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1812-2020
ClassificationClass II
Statusterminated
Initiated2020-03-18
Published2020-05-06
Last updated2026-09-13 11:55 UTC
CompanyWelch Allyn Mortara
DistributionUS

Product

H12+ and patient cable for the H12+ Holter Recorders with the following device name: PAT CBL 10WIRE LEADFORM AHA SNAP, PAT CBL 10WIRE LEADFORM IEC SNAP, PAT CBL 10WIRE LEADFORM XL AHA SNAP, PAT CBL 10WIRE LEADFORM XL IEC SNAP. The 24-hour H12+ recorder uses a single AA alkaline battery to provide continuous 12-lead data recorded over a 24-hour period and a removable 24-hour compact flash (CF) card for data storage.

Reason

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Identifiers

code_infoModel Numbers: H12PLUS-XXX-XXXXX, 9293-017-50, 9293-017-51, 9293-026-50, 9293-026-51.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1812-2020%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.