FDA Device recall Z-1808-2020

ProACT Implantation Instrument Set, is a reusable, steam sterilizable surgical instrument set consisting of a U-Channel Sheath through whic…

On 2020-05-06 the FDA classified a Class II recall of ProACT Implantation Instrument Set, is a reusable, steam sterilizable surgical instrument set consisting of a U-Channel Sheath through whic… by Uromedica, citing potential issue with the manufacture and inspection of U-channel sheaths in Implantation Instruments Sets. Affected U-channel sheaths may be more likely to cause damage to ProACT….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1808-2020
ClassificationClass II
Statusterminated
Initiated2019-04-10
Published2020-05-06
Last updated2026-09-13 11:55 UTC
CompanyUromedica
DistributionUS

Product

ProACT Implantation Instrument Set, is a reusable, steam sterilizable surgical instrument set consisting of a U-Channel Sheath through which the Pro ACT balloon implant is introduced.

Reason

Potential issue with the manufacture and inspection of U-channel sheaths in Implantation Instruments Sets. Affected U-channel sheaths may be more likely to cause damage to ProACT implant devices when passes through the sheaths.

Identifiers

code_infoModel 750041, Lot # UM00650

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.