FDA Device recall Z-1808-2013
Stryker brand Neuroform 3 Microdelivery Stent System, 4.5 mm X 20 mm, REF Catalog Number: E345020, Model/Product Number: M003450200, Produc…
- FDA Device
- Class II
- terminated
- Published 2013-07-31
On 2013-07-31 the FDA classified a Class II recall of Stryker brand Neuroform 3 Microdelivery Stent System, 4.5 mm X 20 mm, REF Catalog Number: E345020, Model/Product Number: M003450200, Produc… by Stryker Neurovascular, citing the device is labeled incorrectly, in that, the device Model Number on the outer carton and inner pouch labels did not match, although both referenced the same lot number. The Mod….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1808-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-06-24 |
| Published | 2013-07-31 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Stryker Neurovascular |
| Distribution | n/a |
Product
Stryker brand Neuroform 3 Microdelivery Stent System, 4.5 mm X 20 mm, REF Catalog Number: E345020, Model/Product Number: M003450200, Product is manufactured by Boston Scientific Corporation, Fremont, CA and distributed by Stryker Neurovascular, Fremont, CA The Neuroform Microdelivery Stent System is intended for use with embolic coils for the treatment of wide neck, intracranial, saccular aneurysms arising from a parent vessel with a diameter of greater than equal to 2 mm and greater than equal to 4.5 mm that are not amenable to treatment with surgical clipping.
Reason
The device is labeled incorrectly, in that, the device Model Number on the outer carton and inner pouch labels did not match, although both referenced the same lot number. The Model Number on the outer box states UPN M003E345020 (4.5mm x 20 mm Neuroform 3 Stent) while the pouch label states UPN M003E3450300 (4.5mm x 30 mm Neuroform 3 Stent).
Identifiers
| code_info | Code: M003450200, Lot 15391019, Expiration Date: 2015-07 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1808-2013%22