FDA Device recall Z-1801-2017

VIDAS Estradiol II, Ref 30431-01, 60 tests The VIDAS Estradiol II (E2 II) assay is intended for use on the instruments of the VIDAS family…

On 2017-04-19 the FDA classified a Class II recall of VIDAS Estradiol II, Ref 30431-01, 60 tests The VIDAS Estradiol II (E2 II) assay is intended for use on the instruments of the VIDAS family… by Biomerieux, citing fulvestrant cross reacts with the VIDAS Estradiol II immunoassay and may cause falsely elevated estradiol results in patients treated with Fulvestrant.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1801-2017
ClassificationClass II
Statusterminated
Initiated2017-03-17
Published2017-04-19
Last updated2026-09-13 11:54 UTC
CompanyBiomerieux
DistributionUS

Product

VIDAS Estradiol II, Ref 30431-01, 60 tests The VIDAS Estradiol II (E2 II) assay is intended for use on the instruments of the VIDAS family (VITEK ImmunoDiagnostic Assay System) as an automated quantitative enzyme-linked fluorescent immunoassay (ELFA) to the determination of total estradiol concentration in human serum or plasma (heparin).

Reason

Fulvestrant cross reacts with the VIDAS Estradiol II immunoassay and may cause falsely elevated estradiol results in patients treated with Fulvestrant.

Identifiers

code_infoLots 1005081700, 1005172850 & 1005359170