RecallRadar

FDA Device recall Z-1798-2021

JELCO HYPODERMIC NEEDLE-PRO FIXED NEEDLE INSULIN SYRINGE: a) 1.0mL 28G X 1/2" (0.33mm x 13mm), REF 4428-1 b) 1.0mL 29G X 1/2" (0.33mm x 13m…

On 2021-06-16 the FDA classified a Class I recall of JELCO HYPODERMIC NEEDLE-PRO FIXED NEEDLE INSULIN SYRINGE: a) 1.0mL 28G X 1/2" (0.33mm x 13mm), REF 4428-1 b) 1.0mL 29G X 1/2" (0.33mm x 13m… by Smiths Medical Asd, citing specific models and lot numbers of the Jelco Hypodermic Needle-Pro Fixed Needle Insulin Syringe 29Gx1/2 odd number line graduation markings on the syringe barrel were skewed appro….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1798-2021
ClassificationClass I
Statusterminated
Initiated2021-05-17
Published2021-06-16
Last updated2026-09-13 11:56 UTC
CompanySmiths Medical Asd
DistributionUS

Product

JELCO HYPODERMIC NEEDLE-PRO FIXED NEEDLE INSULIN SYRINGE: a) 1.0mL 28G X 1/2" (0.33mm x 13mm), REF 4428-1 b) 1.0mL 29G X 1/2" (0.33mm x 13mm), REF 4429-1

Reason

Specific models and lot numbers of the Jelco Hypodermic Needle-Pro Fixed Needle Insulin Syringe 29Gx1/2 odd number line graduation markings on the syringe barrel were skewed approximately 20 degrees upward.

Identifiers

code_info
a) REF 4428-1, Lot Numbers: 4046543 and 4062235 b) REF 4429-1, Lot Numbers: 4031846, 4031845, 4040734, 4043536, 4046…a) REF 4428-1, Lot Numbers: 4046543 and 4062235 b) REF 4429-1, Lot Numbers: 4031846, 4031845, 4040734, 4043536, 4046545, 4046546, 4062239, 4062240, 4062238, 4062242, 4014096

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1798-2021%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.