RecallRadar

FDA Device recall Z-1793-2012

Columbus Tibia Plateau (component of the Columbus REVISION Knee System) Product Usage - The Columbus Revision Knee System is indicated for…

On 2012-06-20 the FDA classified a Class II recall of Columbus Tibia Plateau (component of the Columbus REVISION Knee System) Product Usage - The Columbus Revision Knee System is indicated for… by Aesculap Implant Systems, citing during an internal quality inspection, the tibia plateau was found to be out of specification. There is a potential for misalignment between the tibia plateau and attachment of th….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1793-2012
ClassificationClass II
Statusterminated
Initiated2012-02-21
Published2012-06-20
Last updated2026-09-13 11:53 UTC
CompanyAesculap Implant Systems
DistributionUS

Product

Columbus Tibia Plateau (component of the Columbus REVISION Knee System) Product Usage - The Columbus Revision Knee System is indicated for use in reconstruction of the diseased knee joint caused by osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, the need to revise failed arthroplastics or osteotomies where pain, deformity or dysfunction persist, and for patients suffering from correctable valgus or varus deformity and moderate flexion contracture

Reason

During an internal quality inspection, the tibia plateau was found to be out of specification. There is a potential for misalignment between the tibia plateau and attachment of the extension stem.

Identifiers

code_info
NR068K, TIBIA PLATEAU, FIXED, T0; NR070K, TIBIA PLATEAU, FIXED, T0+; NR072K, TIBIA PLATEAU, FIXED, T1; NR072K, TIBIA PL…NR068K, TIBIA PLATEAU, FIXED, T0; NR070K, TIBIA PLATEAU, FIXED, T0+; NR072K, TIBIA PLATEAU, FIXED, T1; NR072K, TIBIA PLATEAU, FIXED, T1+; NR073K, TIBIA PLATEAU, FIXED, T2; NR074K, TIBIA PLATEAU, FIXED, T2+; NR075K, TIBIA PLATEAU, FIXED, T3; NR076K, TIBIA PLATEAU, FIXED, T3+; NR077K, TIBIA PLATEAU, FIXED, T4; NR078K, TIBIA PLATEAU, FIXED, T4+; NR079K, TIBIA PLATEAU, FIXED, T5