RecallRadar

FDA Device recall Z-1792-2014

cobas 8000 analyzer series Operators Manual Roche cobas Filter label: Filter (N) Mat

On 2014-06-25 the FDA classified a Class II recall of cobas 8000 analyzer series Operators Manual Roche cobas Filter label: Filter (N) Mat by Roche Diagnostics Operations, citing under certain conditions loose ProCell/CleanCell aspiration tube filters may cause: consecutive discrepant results; sipper alarm; foam on reservoirs and system reagent alarms.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1792-2014
ClassificationClass II
Statusterminated
Initiated2014-05-21
Published2014-06-25
Last updated2026-09-13 11:53 UTC
CompanyRoche Diagnostics Operations
DistributionUS

Product

cobas 8000 analyzer series Operators Manual Roche cobas Filter label: Filter (N) Mat. No 03149773001, Contents 1, 768-320 Filter (N) Product Usage: Fully automated immunoassay analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids.

Reason

Under certain conditions loose ProCell/CleanCell aspiration tube filters may cause: consecutive discrepant results; sipper alarm; foam on reservoirs and system reagent alarms.

Identifiers

code_info
1125-07,1017-06,1021-03,1021-05,1016-05,1017-01,1124-05,1124-06,1128-02,1132-03,1132-04,1131-08,…1125-07,1017-06,1021-03,1021-05,1016-05,1017-01,1124-05,1124-06,1128-02,1132-03,1132-04,1131-08,1132-01,1132-02,1128-05,1128-06,2374-11,1132-05,1132-10,1137-02,1139-08,1139-10,1140-07,1140-10,1140-08,1146-01,1146-02,1146-03,1139-06,1139-07,1139-09,1142-09,1145-10,1138-10,1139-01,1127-08,1138-08,1146-05,1146-06,1147-04,1156-04,1140-01,1147-05,1149-01,1149-07,1149-08,1167-02,1156-06,1158-09,1156-05,1158-10,1139-02,1139-03,1139-04,1139-05,1160-01,1160-05,1171-07,1171-08,1171-09,1173-10,1156-01,1160-04,1156-07,1156-02,1156-09,1160-06,1171-10,1173-01,1169-06,1169-08,1169-09,1172-09,1182-08,1211-15,1211-14,1174-08,1160-08,1171-06,1174-06,1174-07,1211-13,1211-06,1183-04,1217-09,1225-11,1211-16,1211-17,1211-18,1224-10,1156-08,1156-10,1225-14,1331-13,1338-15,1338-16,1211-19,1211-04,1211-05,1211-20,

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1792-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.