RecallRadar

FDA Device recall Z-1791-2023

ICD-VR VISIA AF XT OUS IS1/DF1, Model Number DVAB2D1

On 2023-06-28 the FDA classified a Class I recall of ICD-VR VISIA AF XT OUS IS1/DF1, Model Number DVAB2D1 by Medtronic Cardiac Rhythm And Heart Failure Crhf, citing there is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynch….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1791-2023
ClassificationClass I
Statusongoing
Initiated2023-05-10
Published2023-06-28
Last updated2026-09-13 11:56 UTC
CompanyMedtronic Cardiac Rhythm And Heart Failure Crhf
DistributionUS

Product

ICD-VR VISIA AF XT OUS IS1/DF1, Model Number DVAB2D1; Implantable Cardioverter Defibrillators

Reason

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Identifiers

code_info
GTIN 00643169720756, Lot Serial Numbers: BWR600821S, BWR600947S, BWR600936S, BWR600937S, BWR600941S, BWR600946S, BWR60…GTIN 00643169720756, Lot Serial Numbers: BWR600821S, BWR600947S, BWR600936S, BWR600937S, BWR600941S, BWR600946S, BWR600959S, BWR601046S, BWR600938S, BWR601096S, BWR601093S, BWR600979S, BWR601092S, BWR600997S, BWR600934S, BWR600965S, BWR601050S, BWR601049S, BWR600823S, BWR601053S, BWR600983S, BWR600989S, BWR600984S, BWR601055S, BWR601057S, BWR601098S, BWR600822S, BWR600980S, BWR600981S, BWR600982S, BWR601097S, BWR601100S, BWR601095S, BWR601101S, BWR600991S, BWR601051S, BWR601058S, BWR601099S, BWR601094S; GTIN BWR601149S, BWR601155S, BWR601137S, BWR601138S, BWR601108S, BWR601113S, BWR601111S, BWR601119S, BWR601144S, BWR601146S, BWR601147S, BWR601148S, BWR601067S, BWR601150S, BWR601152S, BWR601115S, BWR601105S, BWR601112S, BWR601059S, BWR601061S, BWR601062S, BWR601063S, BWR601072S, BWR601073S, BWR601078S, BWR601081S, BWR601090S, BWR601116S, BWR601103S, BWR601075S, BWR601102S, BWR601104S