FDA Device recall Z-1790-2021
The Sarns Temperature Control and Monitor unit (TCM) is a source of temperature-controlled water for blood heat exchangers used in an extra…
- FDA Device
- Class II
- terminated
- Published 2021-06-09
On 2021-06-09 the FDA classified a Class II recall of The Sarns Temperature Control and Monitor unit (TCM) is a source of temperature-controlled water for blood heat exchangers used in an extra… by Terumo Cardiovascular Systems, citing terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1790-2021 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2021-04-16 |
| Published | 2021-06-09 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Terumo Cardiovascular Systems |
| Distribution | n/a |
Product
The Sarns Temperature Control and Monitor unit (TCM) is a source of temperature-controlled water for blood heat exchangers used in an extracorporeal circuit and for blankets to externally heat or cool the patient. The TCM with options will also supply water for cardioplegia, freeze water for an ice supply, monitor temperatures in the patient and extracorporeal circuit, and allow gradient rewarming relative to a venous blood temperature. Device Name / Model Number: TCM II TUV, 115V (P/N 4415), TCM II TUV, 220V (P/N 4416), Catalog Number: 4416, 164940
Reason
Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices.
Identifiers
| code_info | All lot numbers distributed from 05/02/1985 thru 06/10/2015 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1790-2021%22