RecallRadar

FDA Device recall Z-1787-2015

GEMINI TF PET/CT 64 Slice (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging s…

On 2015-06-17 the FDA classified a Class II recall of GEMINI TF PET/CT 64 Slice (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging s… by Philips Medical Systems Cleveland, citing the firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in batter….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1787-2015
ClassificationClass II
Statusterminated
Initiated2015-03-06
Published2015-06-17
Last updated2026-09-13 11:53 UTC
CompanyPhilips Medical Systems Cleveland
DistributionUS

Product

GEMINI TF PET/CT 64 Slice (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.

Reason

The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.

Identifiers

code_info
Model #: 882471; Serial #s: 7037, 7205, 7206, 7209, 7212, 7213, 7214, 7217, 7218, 7219, 7221, 7227, 7229, 7231, 7234,…Model #: 882471; Serial #s: 7037, 7205, 7206, 7209, 7212, 7213, 7214, 7217, 7218, 7219, 7221, 7227, 7229, 7231, 7234, 7238, 7241, 7245, 7251, 7252, 7253, 7254, 7255, 7256, 7257, 7258, 7504, 7506, 7509, 7510, 7512, 7515, 7516, 7519, 7524, 7525, 7535, 7537, 7539, 7542, 7543, 7544, 7548, 7549, 7550, 7552, 7553, 7555, 7556, 7557, 7558, 7560, 7561, 7562, 7563, 7567, 7569, 7572, 7574, 7577, 7578, 7579, 7585, 7587, 7589, 7592, 7594, 7595, 7602, 7606 & 7575.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1787-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.