FDA Device recall Z-1779-2020
Temperature Sensing 100% Silicone Foley Catheter, Part Number 102201100880MD - Product Usage: is a 3 lumen silicone catheter with a sensor…
- FDA Device
- Class II
- terminated
- Published 2020-04-29
On 2020-04-29 the FDA classified a Class II recall of Temperature Sensing 100% Silicone Foley Catheter, Part Number 102201100880MD - Product Usage: is a 3 lumen silicone catheter with a sensor… by Degania Medical Devices Pvt, citing inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the specified lots of temperature sensing catheters.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1779-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-03-17 |
| Published | 2020-04-29 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Degania Medical Devices Pvt |
| Distribution | US |
Product
Temperature Sensing 100% Silicone Foley Catheter, Part Number 102201100880MD - Product Usage: is a 3 lumen silicone catheter with a sensor to measure temperature while draining the urinary bladder.
Reason
Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the specified lots of temperature sensing catheters.
Identifiers
| code_info | Lot Numbers: S18013318 S18010936 S19006220 S19006221 S19006222 S19006349 |
|---|