FDA Device recall Z-1777-2019
Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86662003 Product Usage: Usage: The Getinge 86 Series Medical Washer-Disinfector…
- FDA Device
- Class II
- terminated
- Published 2019-06-19
On 2019-06-19 the FDA classified a Class II recall of Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86662003 Product Usage: Usage: The Getinge 86 Series Medical Washer-Disinfector… by Maquet Cardiovascular Us Sales, citing possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1777-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-05-20 |
| Published | 2019-06-19 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Maquet Cardiovascular Us Sales |
| Distribution | US |
Product
Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86662003 Product Usage: Usage: The Getinge 86 Series Medical Washer-Disinfector is designed for use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.
Reason
Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process.
Identifiers
| code_info | Serail Numbers: WAA088731, WAA089124 , WAA089128, WAA089586 |
|---|