FDA Device recall Z-1777-2019

Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86662003 Product Usage: Usage: The Getinge 86 Series Medical Washer-Disinfector…

On 2019-06-19 the FDA classified a Class II recall of Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86662003 Product Usage: Usage: The Getinge 86 Series Medical Washer-Disinfector… by Maquet Cardiovascular Us Sales, citing possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1777-2019
ClassificationClass II
Statusterminated
Initiated2019-05-20
Published2019-06-19
Last updated2026-09-13 11:55 UTC
CompanyMaquet Cardiovascular Us Sales
DistributionUS

Product

Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86662003 Product Usage: Usage: The Getinge 86 Series Medical Washer-Disinfector is designed for use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.

Reason

Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process.

Identifiers

code_infoSerail Numbers: WAA088731, WAA089124 , WAA089128, WAA089586

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.