FDA Device recall Z-1772-2022

The Randox Serology I Positive Control contains the following analytes in human based plasma: Anti HIV 1/2, Anti HTLV 1/2, Anti HCV, HBsAg,…

On 2022-09-28 the FDA classified a Class II recall of The Randox Serology I Positive Control contains the following analytes in human based plasma: Anti HIV 1/2, Anti HTLV 1/2, Anti HCV, HBsAg,… by Randox Laboratories, citing ***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control will test negative (Non-Reactiv….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1772-2022
ClassificationClass II
Statusterminated
Initiated2022-08-01
Published2022-09-28
Last updated2026-09-13 11:56 UTC
CompanyRandox Laboratories
DistributionUS

Product

The Randox Serology I Positive Control contains the following analytes in human based plasma: Anti HIV 1/2, Anti HTLV 1/2, Anti HCV, HBsAg, Anti HBc, and Anti Treponema pallidum IgG. The product is delivered as a liquid in a kit of 3 identical ready-to-use vials of 5 mL each.

Reason

***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control will test negative (Non-Reactive) for Marker HBsAg resulting in delay in reporting patient results due to the positive control O36SR testing negative (Non-Reactive) for Marker HBsAg

Identifiers

code_infoGTIN: 05055273216509 Lot number: 036SR Exp Date: 28th March 2023