FDA Device recall Z-1772-2015
Ingenuity Core 128 (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems
- FDA Device
- Class II
- terminated
- Published 2015-06-17
On 2015-06-17 the FDA classified a Class II recall of Ingenuity Core 128 (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems by Philips Medical Systems Cleveland, citing the firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in batter….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1772-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-03-06 |
| Published | 2015-06-17 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Philips Medical Systems Cleveland |
| Distribution | US |
Product
Ingenuity Core 128 (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
Reason
The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Identifiers
| code_info | Model #: 728323; Serial #s: 30153, 32002, 32003, 32004, 32005, 32006, 32007, 32009, 32010, 32011, 32013, 32014, 32015,…Model #: 728323; Serial #s: 30153, 32002, 32003, 32004, 32005, 32006, 32007, 32009, 32010, 32011, 32013, 32014, 32015, 32016, 32018, 32021, 32023, 32025, 32026, 32029, 32030, 32031, 32032, 32033, 32034, 32035, 32036, 32037, 32038, 32040, 32041, 32042, 32043, 32044, 32046, 32047, 32048, 32049, 32051, 32052, 32053, 32054, 32057, 32059, 32060, 32062, 32064, 32066, 32068, 32069, 32070, 32072, 32073, 32074, 32077, 32078, 32080, 32083, 32084, 32089, 32091, 32092, 32094, 32095, 32098, 32100, 32101, 32104, 32106, 32107, 32108, 300137, 310067, 320001, 320002, 320003, 320004, 320005, 320006, 320007, 320008, 320009, 320010, 320011, 320013, 320014, 320015, 320018, 320019, 320020, 320021, 320022, 320023, 320024, 320025, 320026, 320027, 320028, 320029, 320030, 320031, 320032, 320033, 320034, 320035, 320036, 320037, 320038, 320039, 320040, 320041, 320043, 320044, 320045, 320046, 320047, 320048, 320049, 320050, 320051, 320052, 320053, 320054, 320055, 320056, 320057, 320059, 320060, 320062, 320063, 320 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1772-2015%22