Ingenuity Flex (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems

On 2015-06-17 the FDA classified a Class II recall of Ingenuity Flex (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems by Philips Medical Systems Cleveland, citing the firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in batter….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1770-2015
ClassificationClass II
Statusterminated
Initiated2015-03-06
Published2015-06-17
Last updated2026-09-13 11:53 UTC
CompanyPhilips Medical Systems Cleveland
DistributionUS

Product

Ingenuity Flex (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.

Reason

The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.

Identifiers

code_info
Model #: 728317; Serial #s: 260002, 260003, 340002, 340003, 345001, 345002, 345003, 345004, 345005, 345006, 345007, 345…Model #: 728317; Serial #s: 260002, 260003, 340002, 340003, 345001, 345002, 345003, 345004, 345005, 345006, 345007, 345010, 345012, 345013, 345014, 345016, 345017, 345018, 345022, 345023, 345024, 345028 & 345029.