RecallRadar

FDA Device recall Z-1766-2017

The PreludeSYNC Radial Compression Device is a sterile, single use disposable device used to assist in gaining and maintaining hemostasis o…

On 2017-04-19 the FDA classified a Class II recall of The PreludeSYNC Radial Compression Device is a sterile, single use disposable device used to assist in gaining and maintaining hemostasis o… by Merit Medical Systems, citing merit Medical Systems, Inc. announces a voluntary field action for the PreludeSYNC Radial Compression Device because particulate generated during the manufacturing process may ent….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1766-2017
ClassificationClass II
Statusterminated
Initiated2017-03-03
Published2017-04-19
Last updated2026-09-13 11:54 UTC
CompanyMerit Medical Systems
DistributionUS

Product

The PreludeSYNC Radial Compression Device is a sterile, single use disposable device used to assist in gaining and maintaining hemostasis of the radial and ulnar artery following catheterization procedures. It has a soft wristband with a secure hook and loop fastener. The band delivers adjustable compression of the puncture site with an inflatable bulb, and a check valve for easy inflation and deflation with a syringe. A clear curved backer plate provides optimal visualization of the puncture site and ease of placement.

Reason

Merit Medical Systems, Inc. announces a voluntary field action for the PreludeSYNC Radial Compression Device because particulate generated during the manufacturing process may enter the valve, preventing a complete seal of the inflation valve.

Identifiers

code_info
Catalog Number: SRB24AC, Lot Numbers: H1068590S1, H1078177, H1078180, H1083628; Catalog Number: SRB24MED, Lot Numbers:…Catalog Number: SRB24AC, Lot Numbers: H1068590S1, H1078177, H1078180, H1083628; Catalog Number: SRB24MED, Lot Numbers: H1068587S1,H1073974, H1078179, H1078181, H1083629, H1083630; Catalog Number: SRB29AC, Lot Numbers: H1068591S1, H1078183, H1083631; Catalog Number: SRB29MED, Lot Numbers: H1068589S1, H1078184, H1078185, H1083633.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1766-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.