FDA Device recall Z-1760-2017
HYFRECATOR 2000 HANDPIECE SHEATH, REF/Part Number 7-796-19, Rx ONLY, STERILE, Single Patient Use -- CONMED CORPORATION, 525 FRENCH ROAD UTI…
- FDA Device
- Class II
- terminated
- Published 2017-04-19
On 2017-04-19 the FDA classified a Class II recall of HYFRECATOR 2000 HANDPIECE SHEATH, REF/Part Number 7-796-19, Rx ONLY, STERILE, Single Patient Use -- CONMED CORPORATION, 525 FRENCH ROAD UTI… by Conmed, citing for over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches th….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1760-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-02-28 |
| Published | 2017-04-19 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Conmed |
| Distribution | US |
Product
HYFRECATOR 2000 HANDPIECE SHEATH, REF/Part Number 7-796-19, Rx ONLY, STERILE, Single Patient Use -- CONMED CORPORATION, 525 FRENCH ROAD UTICA, NY 13502. --- There is no problem with the Hyfrecator 2000 Electrosurgical Units or with cases of P/N 7-796-19 sold in cases as P/N 7-796-19BX or P/N 7-796-19CS.
Reason
For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were Sterile.
Identifiers
| code_info | Affected Lot Numbers: 201511234, 201512074, 201512094, 201601064, 201603284 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1760-2017%22