FDA Device recall Z-1759-2024

Clinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum covering 38 commonly used clinic…

On 2024-05-15 the FDA classified a Class II recall of Clinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum covering 38 commonly used clinic… by Randox Laboratories, citing randox Laboratories has identified that Inorganic Phosphate in Calibration Serum Level 3, CAL2351 & CAL10388, is running with a negative bias on RX Series instruments compared….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1759-2024
ClassificationClass II
Statusongoing
Initiated2024-03-05
Published2024-05-15
Last updated2026-09-13 11:57 UTC
CompanyRandox Laboratories
DistributionUS

Product

Clinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum covering 38 commonly used clinical chemistry tests.

Reason

Randox Laboratories has identified that Inorganic Phosphate in Calibration Serum Level 3, CAL2351 & CAL10388, is running with a negative bias on RX Series instruments compared to other methods.

Identifiers

code_infoCatalogue No. CAL2351 & CAL10388; GTIN: 05055273200966; Lot No. 1260UE, 1262UE, 1295UE, 1297UE, 1298UE, 1315UE, 1325UE, 1326UE, 1295UE, & 1325UE.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.