FDA Device recall Z-1752-2026
Paradigm REAL-Time Veo Insulin Pump (MMT-554, MMT-754)
- FDA Device
- Class II
- ongoing
- Published 2026-04-08
On 2026-04-08 the FDA classified a Class II recall of Paradigm REAL-Time Veo Insulin Pump (MMT-554, MMT-754) by Medtronic Minimed, citing all Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1752-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-02-13 |
| Published | 2026-04-08 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Medtronic Minimed |
| Distribution | US |
Product
Paradigm REAL-Time Veo Insulin Pump (MMT-554, MMT-754)
Reason
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Identifiers
| code_info | Pump: UDI-DI(GTIN): User Guide Name(Part-Version): Paradigm Veo: 00613994717801, 00613994713834, 00613994713841, 00613…Pump: UDI-DI(GTIN): User Guide Name(Part-Version): Paradigm Veo: 00613994717801, 00613994713834, 00613994713841, 00613994870414, 00613994870377, 00613994870407, 00643169378636, 00613994924957, 00613994717672, 00613994715531, 00613994715555, 00613994715562, 00613994409423, 00613994951502, 00643169026131, 00613994622600, 00613994407740, 00643169077737, 00643169464537, 00643169305038, 00643169500259, 00643169503229, 00643169528789, 00643169533776, 00613994151766, 00643169711440, 00763000084370, 00763000161224, 00763000161262, 00643169464520, 00613994963376, 00643169026100, 00613994924247, 00643169299757, 00643169400931, 00643169400986, 00643169535749, 00643169535794, 00643169519190, 00643169579507, 00643169519091, 00763000084349, 00763000084356, 00763000161248, 00763000048853, 00763000161231, 00643169729445, 00643169846036, 00643169579453, 00613994714183, 00613994716156, 00613994657671, 00643169515741, 00643169579514, 00643169579538, 00643169579477, 00643169579569, 00643169579392, 006431 |
|---|
Status history