RecallRadar

FDA Device recall Z-1751-2026

MiniMed 530G Insulin Pump (MMT-551, MMT-751)

On 2026-04-08 the FDA classified a Class II recall of MiniMed 530G Insulin Pump (MMT-551, MMT-751) by Medtronic Minimed, citing all Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1751-2026
ClassificationClass II
Statusongoing
Initiated2026-02-13
Published2026-04-08
Last updated2026-09-13 11:58 UTC
CompanyMedtronic Minimed
DistributionUS

Product

MiniMed 530G Insulin Pump (MMT-551, MMT-751)

Reason

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

Identifiers

code_info
Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 530G: 00643169507784, 00643169736061, 00643169736092, 00643…Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 530G: 00643169507784, 00643169736061, 00643169736092, 00643169736085, 00643169933224, 00763000161170: Artwork, CRB A Process, Instr, 530G, One Press, Launch, US(MP6026081-012DOC); Artwork, CRB A Process, Instr, 530G, FDA comments, add TST(MP6025813-013DOC); Artwork, CRB A Process, Instr, 530G, Remove Velcro, Icons, US(MP6025813-014DOC);

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1751-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.