RecallRadar

FDA Device recall Z-1750-2022

Surpria 64: Software Version V3.11, V3.22

On 2022-09-21 the FDA classified a Class II recall of Surpria 64: Software Version V3.11, V3.22 by Fujifilm Healthcare Americas, citing error 00003050 or 00003052 may occur during scanning, the captured images will not be reconstructed and the Raw Data will not be displayed in the work list.result, may result in t….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1750-2022
ClassificationClass II
Statusongoing
Initiated2022-07-12
Published2022-09-21
Last updated2026-09-13 11:56 UTC
CompanyFujifilm Healthcare Americas
DistributionUS

Product

Surpria 64: Software Version V3.11, V3.22

Reason

Error 00003050 or 00003052 may occur during scanning, the captured images will not be reconstructed and the Raw Data will not be displayed in the work list.result, may result in the need for additional scanning to capture the images properly for reconstruction, thereby exposing the patient to additional radiation

Identifiers

code_info
UDI: (01)04573596200520(11)220215(21)W5086(250)G1Z0211021 (01)04580292766601(11)200130(21)W8013(250)G1Z0098019 (01)04…UDI: (01)04573596200520(11)220215(21)W5086(250)G1Z0211021 (01)04580292766601(11)200130(21)W8013(250)G1Z0098019 (01)04580292766601(11)200226(21)W8014(250)G1Z0130719 (01)04580292766601(11)201217(21)W8015(250)G1Z0024020 (01)04580292766601(11)210217(21)W8016(250)G1Z0216720 (01)04580292766601(11)210528(21)W8017(250)G1Z0239620 (01)04580292766601(11)210610(21)W8018(250)G1Z0039321 (01)04580292766601(11)210628(21)W8019(250)G1Z0223420 (01)04573596200520(11)210727(21)W8020(250)G1Z0087821 (01)04573596200520(11)210825(21)W8021(250)G1Z0139021 (01)04573596200520(11)211012(21)W8022(250)G1Z0112721 (01)04573596200520(11)211026(21)W8023(250)G1Z0156921 (01)04573596200520(11)220113(21)W8024(250)G1Z0043121 (01)04573596200520(11)220113(21)W8025(250)G1Z0240221 (01)04573596200520(11)220204(21)W8026(250)G1Z0244021 (01)04573596200520(11)220204(21)W8027(250)G1Z0249821 (01)04573596200520(11)220214(21)W8028(250)G1Z0249921 Serial Numbers: G1Z0098019 G1Z0130719 G1Z0024020 G1Z0216720 G1Z0239620 G1Z0

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1750-2022%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.