FDA Device recall Z-1750-2018
Medtronic ZEVO(TM) Anterior Cervical Plate System SLOT SCREW, REF 7710015, Size 4.5mm x 15mm Product Appliance, fixation, spinal interverte…
- FDA Device
- Class II
- terminated
- Published 2018-05-16
On 2018-05-16 the FDA classified a Class II recall of Medtronic ZEVO(TM) Anterior Cervical Plate System SLOT SCREW, REF 7710015, Size 4.5mm x 15mm Product Appliance, fixation, spinal interverte… by Medtronic Sofamor Danek Usa, citing slot screw had an overall length of 15mm, but per the specification, the screw should have an overall length of approximately 17mm.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1750-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-03-23 |
| Published | 2018-05-16 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Medtronic Sofamor Danek Usa |
| Distribution | US |
Product
Medtronic ZEVO(TM) Anterior Cervical Plate System SLOT SCREW, REF 7710015, Size 4.5mm x 15mm Product Appliance, fixation, spinal intervertebral body
Reason
Slot screw had an overall length of 15mm, but per the specification, the screw should have an overall length of approximately 17mm.
Identifiers
| code_info | GTIN/UPN: 643169329218 Lot Number: H5399679 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1750-2018%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.