FDA Device recall Z-1750-2015
Astral 100 ( Model No. 27001) & Astral 150 (Model No. 27003) ventilator
- FDA Device
- Class II
- terminated
- Published 2015-06-17
On 2015-06-17 the FDA classified a Class II recall of Astral 100 ( Model No. 27001) & Astral 150 (Model No. 27003) ventilator by Resmed, citing the Astral device allows clinicians to disable all alarms including those that detect circuit disconnection. Inappropriate disabling of alarms in ventilator dependent patients can….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1750-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-05-05 |
| Published | 2015-06-17 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Resmed |
| Distribution | US |
Product
Astral 100 ( Model No. 27001) & Astral 150 (Model No. 27003) ventilator.
Reason
The Astral device allows clinicians to disable all alarms including those that detect circuit disconnection. Inappropriate disabling of alarms in ventilator dependent patients can lead to hazards of insufficient ventilation and insufficient expiratory pressure which can result in major severity harm.
Identifiers
| code_info | Serial No. 22141256746 22141256747 22141256749 22141256750 22141256752 22141256753 22141256754 22141256755 221…Serial No. 22141256746 22141256747 22141256749 22141256750 22141256752 22141256753 22141256754 22141256755 22141256756 22141256757 22141256758 22141256759 22141256760 22141256761 22141256762 22141256763 22141256764 22141256765 22141256766 22141256767 22141256768 22141256770 22141256771 22141256773 22141256774 22141256776 22141256777 22141256778 22141256780 22141256781 22141256782 22141256783 22141256784 22141256785 22141256786 22141256787 22141256788 22141256789 22141256790 22141256792 22141256793 22141256795 22141256796 22141256797 22141256798 22141256800 22141256802 22141256803 22141256804 22141256805 22141256808 22141256809 22141256810 22141256811 22141256812 22141256813 22141269796 22141269797 22141269798 22141269800 22141269801 22141269802 22141269803 22141269804 22141269805 22141269806 22141269807 22141269808 22141269809 22141269810 22141269811 22141269812 22141269813 22141269814 22141269815 22141269817 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Other recalls by Resmed
- Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients…
FDA Device · Class I · 2026-08-05 · Resmed - AirFit N10 Nasal Masks and User Guide
FDA Device · Class I · 2023-12-27 · Resmed - AirFit F30 Full Face Mask and User Guide
FDA Device · Class I · 2023-12-27 · Resmed - AirFit N20 Nasal Mask and User Guide
FDA Device · Class I · 2023-12-27 · Resmed - AirFit F30i Full Face Mask and User Guide
FDA Device · Class I · 2023-12-27 · Resmed
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1750-2015%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.