FDA Device recall Z-1750-2013

Siemens AXIOM Artis zee/zeego systems equipped with generator A100 Plus Product Usage: Angiographic x-ray system

On 2013-07-24 the FDA classified a Class II recall of Siemens AXIOM Artis zee/zeego systems equipped with generator A100 Plus Product Usage: Angiographic x-ray system by Siemens Medical Solutions Usa, citing during the course of product monitoring, Siemens became aware of a tolerance problem in the power supply of the generator control above a specific value, thus, the generator may f….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1750-2013
ClassificationClass II
Statusterminated
Initiated2013-01-31
Published2013-07-24
Last updated2026-09-13 11:53 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

Siemens AXIOM Artis zee/zeego systems equipped with generator A100 Plus Product Usage: Angiographic x-ray system

Reason

During the course of product monitoring, Siemens became aware of a tolerance problem in the power supply of the generator control above a specific value, thus, the generator may fail.

Identifiers

code_info
Material number 10094137 -- serial numbers 147515, 147500, 147528, 147514, 147516, 147526; material number 10094141 --…Material number 10094137 -- serial numbers 147515, 147500, 147528, 147514, 147516, 147526; material number 10094141 -- serial numbers 154408, 154406, 154453, 154423, 154430, 154425, 154429; material number 10280959 -- serial numbers 160813, 160823, 160812, 160800, 160431, 160814, 160807, 160817, 160816; material number 10094139 -- serial number 157907.