FDA Device recall Z-1747-2026
MiniMed 620G Insulin Pump (MMT-1750)
- FDA Device
- Class II
- ongoing
- Published 2026-04-08
On 2026-04-08 the FDA classified a Class II recall of MiniMed 620G Insulin Pump (MMT-1750) by Medtronic Minimed, citing all Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1747-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-02-13 |
| Published | 2026-04-08 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Medtronic Minimed |
| Distribution | US |
Product
MiniMed 620G Insulin Pump (MMT-1750)
Reason
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Identifiers
| code_info | Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 620G Insulin Pump: 643169559745, 643169721319, 763000375461…Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 620G Insulin Pump: 643169559745, 643169721319, 763000375461, 763000253189: Instr, Startup Guide, 620G 4.10, EN(6025951-022 1); INSTR., NGP 620G, BOLUS WIZARD, EN(6026006-023 1); Instr., 620G, Sftwr. 4.10, EN.(6026006-024 1); INSTRUCTIONS MMT-746EN PUMP 1510/1710 EN(7005933-023 1); DOC PKG MMT-746EN NGP 1710 EN(7005933-024 1); Startup Guide Assembly MMT-755EN(620G(M982438A021 1). |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1747-2026%22