FDA Device recall Z-1747-2023

CRT-D DTBB2Q1 VIVA QUAD S IS4/DF1 INTL, Model Number DTBB2Q1

On 2023-06-28 the FDA classified a Class I recall of CRT-D DTBB2Q1 VIVA QUAD S IS4/DF1 INTL, Model Number DTBB2Q1 by Medtronic Cardiac Rhythm And Heart Failure Crhf, citing there is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynch….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1747-2023
ClassificationClass I
Statusongoing
Initiated2023-05-10
Published2023-06-28
Last updated2026-09-13 11:56 UTC
CompanyMedtronic Cardiac Rhythm And Heart Failure Crhf
DistributionUS

Product

CRT-D DTBB2Q1 VIVA QUAD S IS4/DF1 INTL, Model Number DTBB2Q1; Implantable Cardioverter Defibrillators

Reason

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Identifiers

code_infoGTIN 00643169980181, Lot Serial Numbers: BLQ600039S, BLQ600033S, BLQ600034S, BLQ600035S, BLQ600036S, BLQ600037S, BLQ600038S

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.