RecallRadar

FDA Device recall Z-1744-2026

MiniMed 700G Insulin Pump (MMT-1801, MMT-1805, MMT-1850, MMT-1851)

On 2026-04-08 the FDA classified a Class II recall of MiniMed 700G Insulin Pump (MMT-1801, MMT-1805, MMT-1850, MMT-1851) by Medtronic Minimed, citing all Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1744-2026
ClassificationClass II
Statusongoing
Initiated2026-02-13
Published2026-04-08
Last updated2026-09-13 11:58 UTC
CompanyMedtronic Minimed
DistributionUS

Product

MiniMed 700G Insulin Pump (MMT-1801, MMT-1805, MMT-1850, MMT-1851)

Reason

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

Identifiers

code_info
Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 700G Insulin pump: 763000552541, 763000541323, 763000854737…Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 700G Insulin pump: 763000552541, 763000541323, 763000854737, 763000854720: MINIMEDTM 700 INSULIN PUMP USER GUIDE mmol/L(M984972A021 1); MINIMED 700 INSULIN PUMP USER GUIDE mg/dL(M015850C001 1).

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1744-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.