RecallRadar

FDA Device recall Z-1741-2013

CORE Universal Driver REF 5400- 99 Rx Only This drill may also be used with theTotal Performance System (TPS").The Stryker Total Performanc…

On 2013-07-24 the FDA classified a Class II recall of CORE Universal Driver REF 5400- 99 Rx Only This drill may also be used with theTotal Performance System (TPS").The Stryker Total Performanc… by Stryker Instruments Div, citing the safety margin values detailed in the Engineering Design were entered incorrectly into the programming software. Handpieces that have been programmed with the incorrect paramet….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1741-2013
ClassificationClass II
Statusterminated
Initiated2013-07-01
Published2013-07-24
Last updated2026-09-13 11:53 UTC
CompanyStryker Instruments Div
DistributionUS

Product

CORE Universal Driver REF 5400- 99 Rx Only This drill may also be used with theTotal Performance System (TPS").The Stryker Total Performance System (TPS) is intended for usein the cutting, drilling, reaming, decorticating and smoothing of bone and other bone related tissues for a variety of applications such as ENT, dental orthopedic, maxillofacial, spinal, and plastic surgery. The instruments are also used in the placement of screws, wires, pins, and other fixation devices.

Reason

The safety margin values detailed in the Engineering Design were entered incorrectly into the programming software. Handpieces that have been programmed with the incorrect parameters could result in unintended activation. There is a potential for injury to the patient or user as a result of unintended forward or reverse activation. There is also the potential that the handpiece may not reach 100

Identifiers

code_info
Serial Number 0326801633, 0330802913, 0406200763, 0410401383, 0410600013, 0534203093,0627600783, 0730101013, 0730101033…Serial Number 0326801633, 0330802913, 0406200763, 0410401383, 0410600013, 0534203093,0627600783, 0730101013, 0730101033, 0805000553, 0809829593, 0935603083,1011903103, 1011903493, 1013814003, 1025400463, 1025400643, 1126901343, and 1128022593