RecallRadar

FDA Device recall Z-1737-2022

Disposable Subdermal Needle Electrode, Pt/lr, 12 x 0.40mm, product number TE/S46/638, 25 needles per package, sterile

On 2022-09-21 the FDA classified a Class II recall of Disposable Subdermal Needle Electrode, Pt/lr, 12 x 0.40mm, product number TE/S46/638, 25 needles per package, sterile by Technomed Europe, citing burns of varying degrees can occur to patients with electrodes left in while in the MRI scanner as a result of inadequate safety information in the IFU.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1737-2022
ClassificationClass II
Statusterminated
Initiated2022-08-09
Published2022-09-21
Last updated2026-09-13 11:56 UTC
CompanyTechnomed Europe
DistributionUS

Product

Disposable Subdermal Needle Electrode, Pt/lr, 12 x 0.40mm, product number TE/S46/638, 25 needles per package, sterile.

Reason

Burns of varying degrees can occur to patients with electrodes left in while in the MRI scanner as a result of inadequate safety information in the IFU.

Identifiers

code_infoLot numbers 044220, exp. 1/1/2023; 045077, exp. 3/1/2023; 045443, exp. 4/1/2023; and 049285, exp. 7/1/2024; UDI-DI 08718375861530.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1737-2022%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.