FDA Device recall Z-1734-2019
GETINGE MEERA MOBILE OPERATING TABLE: (1) Part 720001B2 (2) Part 720001B0 (3) Part 720001F2 (4) Part 720001F0 (5) Part 710001B2 (6) Part 71…
- FDA Device
- Class II
- terminated
- Published 2019-06-12
On 2019-06-12 the FDA classified a Class II recall of GETINGE MEERA MOBILE OPERATING TABLE: (1) Part 720001B2 (2) Part 720001B0 (3) Part 720001F2 (4) Part 720001F0 (5) Part 710001B2 (6) Part 71… by Maquet Cardiovascular Us Sales, citing user may suffer an electric shock when coming in contact with the exposed electronic components.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1734-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-05-06 |
| Published | 2019-06-12 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Maquet Cardiovascular Us Sales |
| Distribution | US |
Product
GETINGE MEERA MOBILE OPERATING TABLE: (1) Part 720001B2 (2) Part 720001B0 (3) Part 720001F2 (4) Part 720001F0 (5) Part 710001B2 (6) Part 710001B0 (7) Part 710001F0 Product Usage: The Meera Mobile Operating Table is designed to support and position the patient immediately before, during and after surgical interventions as well as for examination and treatment.
Reason
User may suffer an electric shock when coming in contact with the exposed electronic components
Identifiers
| code_info | (1) Part 720001B2 S/N 1 to S/N 416 (2) Part 720001B0 S/N 1 to S/N 1416 (3) Part 720001F2 S/N 1 to S/N 91 (4…(1) Part 720001B2 S/N 1 to S/N 416 (2) Part 720001B0 S/N 1 to S/N 1416 (3) Part 720001F2 S/N 1 to S/N 91 (4) Part 720001F0 S/N 1 to S/N 220 (5) Part 710001B2 S/N 1 to S/N 8 (6) Part 710001B0 S//N 1 to S/N 11 (7) Part 710001F0 S/N 1 Expanded lots: 720001B2 SN 417 to SN 503 720001B0 SN 1417 to SN 1689 720001F2 SN 92 to SN 119 720001F0 SN 221 to SN 226 and SN 701 to 711 710001B2 SN 9 to SN 14 710001B0 SN 12 to SN 88 |
|---|