FDA Device recall Z-1729-2026
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as foll…
- FDA Device
- Class I
- ongoing
- Published 2026-04-15
On 2026-04-15 the FDA classified a Class I recall of Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as foll… by Medline Industries, citing medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connectio….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1729-2026 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2026-02-27 |
| Published | 2026-04-15 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Medline Industries |
| Distribution | US |
Product
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: TAVR PACK, Medline SKU # DYNJ84472
Reason
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Identifiers
| code_info | Medline SKU # DYNJ84472, UDI/DI each 10195327369286, UDI/DI case 40195327369287, Lot Numbers: 23LBN016, 23KBL255, 23HBL885, 23FBV799. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1729-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.