Assay Data Disk (ADD), All DRV versions that support product within expiry dating Automates pre-analytical and post-analytical sample and d…

On 2015-06-17 the FDA classified a Class II recall of Assay Data Disk (ADD), All DRV versions that support product within expiry dating Automates pre-analytical and post-analytical sample and d… by Ortho Clinical Diagnostics, citing software anomaly regarding urine samples that require acidified pretreatment. The software has allowed urine samples for assays that require acidified pretreatment to be metered f….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1729-2015
ClassificationClass II
Statusterminated
Initiated2015-04-06
Published2015-06-17
Last updated2026-09-13 11:53 UTC
CompanyOrtho Clinical Diagnostics
DistributionUS

Product

Assay Data Disk (ADD), All DRV versions that support product within expiry dating Automates pre-analytical and post-analytical sample and data management for in vitro diagnostic use. VITROS Urine Assays: VITROS Calcium (Ca), Magnesium (Mg), and Phosphorus (Phos) requires acidification pretreatment prior to urine assay testing

Reason

Software anomaly regarding urine samples that require acidified pretreatment. The software has allowed urine samples for assays that require acidified pretreatment to be metered from the same sample container as assays that required no pretreatment when the sample was programmed.

Identifiers

code_infoProduct Code: 6801876 Unique Device Identifier Numbers: 10758750001576