FDA Device recall Z-1729-2014
HLD Systems Model 610 Washer/High Level Disinfector
- FDA Device
- Class II
- terminated
- Published 2014-06-18
On 2014-06-18 the FDA classified a Class II recall of HLD Systems Model 610 Washer/High Level Disinfector by Cenorin, citing cenorin LLC is voluntarily recalling HLD Systems Model 610 Washer/High Level Disinfector due to a potential risk of the system not injecting the proper amount of cleaning fluid du….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1729-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-04-24 |
| Published | 2014-06-18 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Cenorin |
| Distribution | US |
Product
HLD Systems Model 610 Washer/High Level Disinfector. Maximum weight of the Model 610 when filled with water is approximately 450 pounds. The dimensions of the system are: Width: 26.5" (67 cm); Depth: 33" (83 cm); Height: 46.5" (118 cm) to top of control panel; 64" (162.6 cm)- height clearance with lid open. Product Usage: Usage: This device washes reusable medical devices and disinfects them with high temperature hot water (pasteurization). The devices are washed with a cleaning solution (containing NaClO (sodium hypochlorite), surfactants, and detergents) to aid in the removal of debris and visible stains. Next, the medical devices and rinsed two times with water. The final, disinfection step uses only hot water at 160 F ¿ 2 F for 30 minutes.
Reason
Cenorin LLC is voluntarily recalling HLD Systems Model 610 Washer/High Level Disinfector due to a potential risk of the system not injecting the proper amount of cleaning fluid during the wash cycle and not achieving the desired level of debris reduction.
Identifiers
| code_info | 610601, 610603, 610703, 610704, 610705, 610707, 610708, 610802, 610803, 610804, 610901, 610903, 610904, 610905, 610902,…610601, 610603, 610703, 610704, 610705, 610707, 610708, 610802, 610803, 610804, 610901, 610903, 610904, 610905, 610902, 611001, 611002, 611004, 611102, 611103, 611201, 611202, 611203, 610902L, 611204. First two digits (61) refer to model number 610 for the HLD system Second two digits - two digit year of manufacture Last two digits sequentially numbered as they receive orders. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1729-2014%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.