FDA Device recall Z-1727-2014
Alere Triage¿ D-Dimer Test PN 98100, Lot W53884B. The Alere Triage¿ D-Dimer Test is a fluorescence immunoassay to be used with the Alere Tr…
- FDA Device
- Class II
- terminated
- Published 2014-06-18
On 2014-06-18 the FDA classified a Class II recall of Alere Triage¿ D-Dimer Test PN 98100, Lot W53884B. The Alere Triage¿ D-Dimer Test is a fluorescence immunoassay to be used with the Alere Tr… by Alere San Diego, citing alere initiated this recall because a limited number of Alere Triage¿ D-Dimer devices from PN 98100 Lot W53884B were incorrectly manufactured. As a result, when testing patient sa….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1727-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-05-02 |
| Published | 2014-06-18 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Alere San Diego |
| Distribution | US |
Product
Alere Triage¿ D-Dimer Test PN 98100, Lot W53884B. The Alere Triage¿ D-Dimer Test is a fluorescence immunoassay to be used with the Alere Triage¿ Meters for the quantitative determination of cross-linked fibrin degradation products containing D-dimer in EDTA anticoagulated whole blood and plasma specimens. . The Alere Triage¿ D-Dimer test is intended for use as an aid in the assessment and evaluation of patients suspected of having disseminated intravascular coagulation or thromboembolic events including pulmonary embolism.
Reason
Alere initiated this recall because a limited number of Alere Triage¿ D-Dimer devices from PN 98100 Lot W53884B were incorrectly manufactured. As a result, when testing patient sample these affected devices will either generate an error code (e.g. E4 error), or they may generate an incorrect result.
Identifiers
| code_info | PN 98100, Lot W53884B. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1727-2014%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.