FDA Device recall Z-1727-2013
B-CrossLaps/serum (B-CTx in serum) REF 11972308 160 100 tests Immunoassay for the in vitro quantitative determination of degradation produc…
- FDA Device
- Class II
- terminated
- Published 2013-07-24
On 2013-07-24 the FDA classified a Class II recall of B-CrossLaps/serum (B-CTx in serum) REF 11972308 160 100 tests Immunoassay for the in vitro quantitative determination of degradation produc… by Roche Diagnostics Operations, citing roche Diagnostics internal investigations in R&D have shown that the claims for biotin interference and the sample stability given in the Elecsys ¿-CrossLaps (¿-CL) assay package….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1727-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-04-25 |
| Published | 2013-07-24 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Roche Diagnostics Operations |
| Distribution | US |
Product
B-CrossLaps/serum (B-CTx in serum) REF 11972308 160 100 tests Immunoassay for the in vitro quantitative determination of degradation products of type I collagen in human serum and plasma as an aid in assessing bone resorption. The test may be used as an aid in monitoring antiresorptive therapies (e.g. bisphosphonates, hormone replacement therapy - HRT) in postmenopausal women and individuals diagnosed with osteopenia. The electrochemiluminescence immunoassay ECLIA is intended for use on Elecsys and cobas e immunoassay analyzers.
Reason
Roche Diagnostics internal investigations in R&D have shown that the claims for biotin interference and the sample stability given in the Elecsys ¿-CrossLaps (¿-CL) assay package insert version 5, could not be reproduced. The ¿-CrossLaps test results contribute to the monitoring of antiresorptive therapy. False low or false normal results may lead to delayed diagnosis and treatment of osteoporosi
Identifiers
| code_info | All lot numbers |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.