FDA Device recall Z-1719-2014
MACROLYTE Premie Dispersive Electrode, REF 440-2400, CONMED CORPORATION. The device is applied to the patient during electrosurgical proced…
- FDA Device
- Class II
- terminated
- Published 2014-06-11
On 2014-06-11 the FDA classified a Class II recall of MACROLYTE Premie Dispersive Electrode, REF 440-2400, CONMED CORPORATION. The device is applied to the patient during electrosurgical proced… by Conmed, citing conMed Corporation has received complaints regarding sparking, no output and burning at the cord set connection. The complaints have been confirmed by ConMed.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1719-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-05-08 |
| Published | 2014-06-11 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Conmed |
| Distribution | US |
Product
MACROLYTE Premie Dispersive Electrode, REF 440-2400, CONMED CORPORATION. The device is applied to the patient during electrosurgical procedures to provide a path for RF current to leave the patient and return to the generator.
Reason
ConMed Corporation has received complaints regarding sparking, no output and burning at the cord set connection. The complaints have been confirmed by ConMed.
Identifiers
| code_info | Beginning Lot Code 1204275; Ending Lot Code 1401075 (Lot code indicates the date of manufacture = yymmdds, where s is the shift). |
|---|