Patient Side Cart assembly, 4-arm, for use with Intuitive Surgical, Inc. da Vinci Surgical System IS 4000

On 2014-06-11 the FDA classified a Class II recall of Patient Side Cart assembly, 4-arm, for use with Intuitive Surgical, Inc. da Vinci Surgical System IS 4000 by Intuitive Surgical, citing torque wrenches used to tighten certain PSC bolts were out of calibration. This could lead to over- or under-tightening of bolts that may result in system faults, mechanical sound….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1718-2014
ClassificationClass II
Statusterminated
Initiated2014-05-09
Published2014-06-11
Last updated2026-09-13 11:53 UTC
CompanyIntuitive Surgical
DistributionFL, TX

Product

Patient Side Cart assembly, 4-arm, for use with Intuitive Surgical, Inc. da Vinci Surgical System IS 4000. The patient-side cart is positioned at the operating room table and contains four arms that are positioned over the target patient anatomy. The endoscope attaches onto any arm. The patient-side assistant attaches/detaches the endoscope and instrument intra-operatively.

Reason

Torque wrenches used to tighten certain PSC bolts were out of calibration. This could lead to over- or under-tightening of bolts that may result in system faults, mechanical sound (e.g. scraping, squeaking, rattling), cracks or vibration in the system.

Identifiers

code_infolot 380652-20