FDA Device recall Z-1717-2020
Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers
- FDA Device
- Class II
- terminated
- Published 2020-04-22
On 2020-04-22 the FDA classified a Class II recall of Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers by Sysmex America, citing when excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1717-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-01-21 |
| Published | 2020-04-22 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Sysmex America |
| Distribution | FL, IL, VT |
Product
Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers.
Reason
When excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a combustion hazard.
Identifiers
| code_info | Serial Numbers: 11201, 11202, 12310, 12311, 12312, 12313, 12941, 12942, 12997, 13875, 14170, 14171, 14172, 14173, 1417…Serial Numbers: 11201, 11202, 12310, 12311, 12312, 12313, 12941, 12942, 12997, 13875, 14170, 14171, 14172, 14173, 14174, 14175, 14278, 14279, 14280, 14962, 14963, 14964, 14965, 14966, 14967, 14968, 14969, 14970, 14972, 14973, 14974, 14975, 14976, 14977, 14978, 14979, 15017, 15018, 15619, 15620 |
|---|