FDA Device recall Z-1712-2017

ADVIA Chemistry Enzyme 2 Calibrator, multi-analyte mixture Product Usage: Assays included in Enzyme 2 Calibrator: AST, AST_c ALT, ALT_c, AL…

On 2017-04-12 the FDA classified a Class III recall of ADVIA Chemistry Enzyme 2 Calibrator, multi-analyte mixture Product Usage: Assays included in Enzyme 2 Calibrator: AST, AST_c ALT, ALT_c, AL… by Siemens Healthcare Diagnostics, citing siemens Healthcare Diagnostics has confirmed a positive bias of up to 41% in quality control and patient results for Aspartate Aminotransferase (AST) and Aspartate Aminotransferas….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1712-2017
ClassificationClass III
Statusterminated
Initiated2017-02-23
Published2017-04-12
Last updated2026-09-13 11:54 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

ADVIA Chemistry Enzyme 2 Calibrator, multi-analyte mixture Product Usage: Assays included in Enzyme 2 Calibrator: AST, AST_c ALT, ALT_c, ALTP5P, ALTP_c, ASTP5P and ASTP_c

Reason

Siemens Healthcare Diagnostics has confirmed a positive bias of up to 41% in quality control and patient results for Aspartate Aminotransferase (AST) and Aspartate Aminotransferase concentrated (AST_c) assays following calibration with Advia Chemistry Enzyme 2 Calibrator lots 6JD079 and 6MD042.

Identifiers

code_infoLots 6JD079 and 6MD042