RecallRadar

FDA Device recall Z-1711-2014

AXIOM Luminos dRFsystem The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluorosc…

On 2014-06-11 the FDA classified a Class II recall of AXIOM Luminos dRFsystem The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluorosc… by Siemens Medical Solutions Usa, citing it was determined that under rare environmental conditions (i.e., extremely dry air in combination with a highly isolated floor) electrostatic discharges (ESD) >8 kV may trigger a….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1711-2014
ClassificationClass II
Statusterminated
Initiated2014-04-11
Published2014-06-11
Last updated2026-09-13 11:53 UTC
CompanySiemens Medical Solutions Usa
DistributionCA, CO, FL, GA, IL, IN, ME, MI, MN, MO, NJ, NY, OH, OK, OR, PA, TN, TX, VA

Product

AXIOM Luminos dRFsystem The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image.

Reason

It was determined that under rare environmental conditions (i.e., extremely dry air in combination with a highly isolated floor) electrostatic discharges (ESD) >8 kV may trigger an unintended movement of the AXIOM Luminos dRFsystem which may lead to an emergency situation and pose danger to the patients, operating personnel or to the unit.

Identifiers

code_info
Model number: 10094200, serial numbers:4332 4291 4323 4430 3061 4331 4238 4297 4201 4024 3098 4252 4372 43…Model number: 10094200, serial numbers:4332 4291 4323 4430 3061 4331 4238 4297 4201 4024 3098 4252 4372 4354 4077 3050 3108 4327 4342 4251 3053 3114 4156 3145 3096 4043 3181 4208 3090 4143 4051 4199 4401 4405 4151 4373 4276